Web based remote monitoring
Remote Monitoring with WiFi
Sentinel Next 1S
Remote Monitoring with LTE (cell) Sentinel Next 1S LTE
IoT Practical and Usable
The Sentinel Next App provides all the tools you need to monitor your important assets on the go
• Sensor Configuration Wizard
• Push notification alarms
• Sensor signing
• Graphs/Data
• Sensor Health
IoT Cloud Services
Your data is stored on the Argus Labs Cloud
IoT Cloud-monitoring service designed for the scientific, business and technology needs of the Pharmaceutical, Blood Bank, Biotech, Hospitals, Pharmacy and Life Sciences industries.
The solution is 21 CFR Part 11 compliant.
• Evaluation of our sensors
• Low to large number of sensors
• Setups which require web access
• Monitoring 10+ types of sensors
• New asset tracking feature (call us)
Monitoring Products
Sentinel Next RTD Temperature Probe with Dry Contact
Sentinel Next RTD Temperature Probe with Dry ContactSentinel Next RTD Temperature Probe with Dry Contact
Sentinel Next RTD Temperature Probe with Dry ContactSentinel Next RTD probe with Dry Contact for ultra-low freezers and liquid nitrogen chambers. Used in tissue and blood storage, pharmaceutical and other healthcare applications
Probe Length & Diameter Stainless probe: 25mm, Ø 4mm (1”, Ø 0.16”)
Cable Length: 183cm (6ft)
Measurement Range: -200°C to +150°C (-328°F to +302°F)
Accuracy: ±0.2°C (±0.4°F) at 25°C (77°F)
Response Time: (to reach 90%) In water, with stirring: <10sec
Display Resolution: 0.1°C (0.1°F)
Optional DI (Digital Input): dry contact available
Door Contact
Circuit type Open loop
Operation gap ½”(13mm)
Color: White
Terminals: Screw-type
Mounting Tabs: Yes
Case: High-impact ABS plastic
Contacts: Deactivated rhodium on gold under-plating
Contact rating: 0.1A@100VDC (max) 1.0A@10VDC (max) 10W (max)
Temperature: -15°~160°F (-25°~70°C)
Switch cycles: 50 Million (0.1mA@5VDC)
Switch Dimensions: 21/2′′x9/16′′x1/2′′ (63.5x14x13mm)
Magnet Dimensions: 21/2′′x1/2′′x1/2′′ (63.5x13x13mm)
Magnet Type: Ferrite
Remote Monitoring – Sentinel Next 1S
Remote Monitoring – Sentinel Next 1SRemote Monitoring – Sentinel Next 1S
Remote Monitoring – Sentinel Next 1SUses your existing WiFi network to monitor across all your facilities. Automates monitoring of temperature, humidity, differential pressure, CO2 and more to meet compliance standards with 24/7 real-time continuous monitoring. Alerting system uses emails, text messages and APPs for notifications. The system meets 21 CFR 11 requirements, are ideal for monitoring of refrigerators, freezers and ultralow freezers.
[maxbutton id="1" url="https://www.aegisfridge.com/contact/" ]RTD Probe
RTD ProbeRTD Probe
RTD ProbeRTD Probe for ultra-low freezers and liquid nitrogen chambers. Used in tissue and blood storage, pharmaceutical and other healthcare applications
[maxbutton id="1" ]
Sentinel Next RTD – Ultra Low Temperature Probe
Sentinel Next RTD – Ultra Low Temperature ProbeSentinel Next RTD – Ultra Low Temperature Probe
Sentinel Next RTD – Ultra Low Temperature ProbeSentinel Next Ultra Low Temperature Probe for ultra-low freezers and liquid nitrogen chambers. Used in tissue and blood storage, pharmaceutical and other healthcare applications
Probe Length & Diameter Stainless probe: 25mm, Ø 4mm (1”, Ø 0.16”)
Cable Length: 183cm (6ft)
Measurement Range: -200°C to +150°C (-328°F to +302°F)
Accuracy: ±0.2°C (±0.4°F) at 25°C (77°F)
Response Time: (to reach 90%) In water, with stirring: <10sec
Display Resolution: 0.1°C (0.1°F)
Optional DI (Digital Input): dry contact available
Sentinel Next – Temperature and Relative Humidity Connector
Sentinel Next – Temperature and Relative Humidity ConnectorSentinel Next – Temperature and Relative Humidity Connector
Sentinel Next – Temperature and Relative Humidity ConnectorSentinel Next Temperature and Relative Humidity Connector
Temperature Measurement Range: -40°C to +125°C (-40°F to +257°F)
Accuracy: ±0.3°C (±0.5°F)
Relative Humidity Range: 0 to +100% RH
Response Time: (to reach 66%) 5 to 30sec for T; 8sec for RH
Display Resolution: 0.1°C (0.1°F)
Remote Monitoring – Sentinel Next 1S LTE
Remote Monitoring – Sentinel Next 1S LTERemote Monitoring – Sentinel Next 1S LTE
Remote Monitoring – Sentinel Next 1S LTEUses the cell phone network to monitor across all your facilities thereby eliminating IT security concerns, network performance issues and support commitment from IT. Automates monitoring of temperature, humidity, differential pressure, CO2 and more to meet compliance standards with 24/7 real-time continuous monitoring. Alerting system uses emails, text messages and APPs for notifications. The system meets 21 CFR 11 requirements, are ideal for monitoring of refrigerators, freezers and ultralow freezers.
[maxbutton id="1" url="https://www.aegisfridge.com/contact/" ]Sentinel Next Large Range CO2 Probe (0-30%)
Sentinel Next Large Range CO2 Probe (0-30%)Sentinel Next Large Range CO2 Probe (0-30%)
Sentinel Next Large Range CO2 Probe (0-30%)Probe Dimensions: 7.6cm x 9.5cm x 3.8cm (3” x 3.75” x 1.5”)
Cable Length: 100cm (~39”)
Operating Principle: Non-dispersive infrared (NDIR)
Measurement range: CO2 0 to 30%vol(CO2)
Measurement RH: 0 to 100% (non-condensing)
Measurement range: -40 to 60 °C
Accuracy: ± 0,25vol ± 3% of reading
Operation temperature range: 0 to 50 °C
Operation humidity range: 0 to 95% RH (non-condensing)
Display Resolution: 0.1°C (0.1°F)
[maxbutton id="1" ]
Glycol Buffer Bottle
Glycol Buffer BottleGlycol Buffer Bottle
Glycol Buffer Bottle- Use a thermal buffer with your probe to
prevent nuisance alarms.
Glycol, Glass Beads, IceClear PGX. Specify buffer solution on order.
[maxbutton id="1" ]
Sentinel Next PRO Temperature Probe
Sentinel Next PRO Temperature ProbeSentinel Next PRO Temperature Probe
Sentinel Next PRO Temperature ProbeSentinel Next PRO Probe for ultra-low freezers and liquid nitrogen chambers. Used in tissue and blood storage, pharmaceutical and other healthcare applications
Probe Length & Diameter Stainless probe: 25mm, Ø 4mm (1”, Ø 0.16”)
Cable Length: 304cm (10ft)
Measurement Range: -30°C to +150°C (-22°F to +302°F)
Accuracy: ±0.2°C (±0.4°F) at 25°C (77°F)
Response Time (to reach 90%) In water, with stirring: <10sec
Display Resolution: 0.1°C (0.1°F)
Sentinel Next Differential Pressure, +/- 125 Pa
Sentinel Next Differential Pressure, +/- 125 PaSentinel Next Differential Pressure, +/- 125 Pa
Sentinel Next Differential Pressure, +/- 125 PaOutput: I2C
Pneumatic Connection: Manifold or Tube
Pressure range (bidirectional): 125 Pa 0.5” H2O
Accuracy of measured value: 3%
Lowest detectable pressure: < 0.01 Pa
Measurement speed: 0.5 ms
Calibrated for: Air, N2
Gas compatibility: Air, inert gases
[maxbutton id="1" ]